Journal of Aerosol Science, Vol.28, No.3, 471-482, 1997
Design of an instrument for real-time detection of bioaerosols using simultaneous measurement of particle aerodynamic size and intrinsic fluorescence
A prototype instrument has been constructed to measure individual airborne particles based on their aerodynamic size and their intrinsic fluorescence at selected excitation and emission wavelength bands. The instrument combines features of an aerodynamic particle sizing device with capabilities similar to those of a liquid flow cytometer. The goal of the instrument is to provide real-time data indicative of particle characteristics, and it is especially targeted to respond to bioaerosols from 0.5 to 10 mu m (aerodynamic diameter) with intrinsic fluorescence excited at a wavelength of 325 nm and emitting from 420 to 580 nm. This size range covers individual airborne bacteria and bacterial clusters, and the fluorescence sensitivity is Selected for biological molecules commonly found in cellular systems, for example, reduced nicotinamide adenine dinucleotide phosphate [NAD(P)H] and riboflavin. Initial tests with nebulised Bacillus subtilis var. niger (BG, ATCC 9372) spores have shown that, for both individual spores and spore clumps, a low level of fluorescence is detected from 17% of the particles. This detection percentage is on the same order as previous experiments that have measured viability of about 12% for mechanically dispersed BG spores (Ho and Fisher (1993) Defense Research Establishment Suffield Memorandum 1421) and suggests a need for further investigation into the possible relationship between the detected fluorescence and viability of bacterial spores.
Keywords:BACTERIA