Electrophoresis, Vol.35, No.7, 956-961, 2014
Very fast electrophoretic determination of creatinine and uric acid in human urine using a combination of two capillaries with different internal diameters
A capillary system formed by combining 25 and 100 m id capillaries was used in the short-end injection mode to determine creatinine and uric acid in human urine. The separation was performed at an electric field intensity of 2.3 kV/cm. Creatinine was determined in a BGE with a composition of 20 mM citric acid/NaOH (pH 3.0), and uric acid was determined in 20 mM MES/NaOH (pH 6.0). Under these conditions, migration times of 12.2 s for creatinine and 8.6 s for uric acid were achieved. The LOD value is 2.4 mg/L for creatinine and 0.9 mg/L for uric acid; the RSD for the migration time varies in the range 0.7-1.1% (intra day) to 1.0-7.5% (inter day); RSDs for the peak areas equalled 3.4-4.0% (intra day) and 4.3-4.7% (inter day). The determined creatinine values in seven urine samples vary in the range 221-1394 mg/L for creatinine and 87-615 mg/L for uric acid. t-Test did not reveal any statistically significant difference between the developed CE methodologies and reference methods - Jaffe reaction for creatinine and enzymatic uricase test for uric acid.