화학공학소재연구정보센터
Electrophoresis, Vol.35, No.17, 2534-2537, 2014
Application of a capillary-assembled microfluidic system for separation of cephalosporins
This paper demonstrates a simple and easy setting up of a fused-silica capillary-assembled microfluidic system (CE). This system incorporates a split-flow pressure injection of the sample into a microfluidic system made from PDMS and a short (approximate to 20 cm) length of fused-silica capillary as a separation unit. The on-capillary detection was carried out by fiber optic spectrometry. A mixture of six cephalosporin antibiotics was separated in the CE system and the obtained results were compared to those achievable by conventional CE. The six components could be separated within 8.5 min with the number of theoretical plates around 10000.