화학공학소재연구정보센터
Chemical Physics Letters, Vol.653, 5-10, 2016
A novel method for measuring the concentration of chloroform based on kinetic parameters at atmosphere
A novel method is proposed to detect chloroform concentrations based on the kinetic parameters using ion mobility spectrometer with a negative corona discharge ion source operating at atmospheric pressure. Unlike conventional sample introduction mode, in this technique, CHCl3 enters into the drift tube from the end of drift region carried by the drift gas. There are two tails before Cl- and (CHCl3)center dot Cl- ion peaks, which fit to the ions formed in the drift region. Utilizing the kinetic parameters, concentration for CHCl3 can be calculated. This method not only offers a new way to get concentrations of CHCl3 under atmospheric pressure. (C) 2016 Elsevier B.V. All rights reserved.