Nature Materials, Vol.16, No.3, 303-303, 2017
Instrumented cardiac microphysiological devices via multimaterial three-dimensional printing
Biomedical research has relied on animal studies and conventional cell cultures for decades. Recently, microphysiological systems (MPS), also known as organs-on-chips, that recapitulate the structure and function of native tissues in vitro, have emerged as a promising alternative(1). However, current MPS typically lack integrated sensors and their fabrication requires multi-step lithographic processes(2). Here, we introduce a facile route for fabricating a newclass of instrumented cardiac microphysiological devices via multimaterial three-dimensional(3D) printing. Specifically, we designed six functional inks, based on piezo-resistive, high-conductance, and biocompatible soft materials that enable integration of soft strain gauge sensors within micro-architectures that guide the self-assembly of physio-mimetic laminar cardiac tissues. We validated that these embedded sensors provide non-invasive, electronic readouts of tissue contractile stresses inside cell incubator environments. We further applied these devices to study drug responses, as well as the contractile development of human stem cell-derived laminar cardiac tissues over four weeks.