화학공학소재연구정보센터
Electrochimica Acta, Vol.245, 319-328, 2017
Tailoring the electrochemical degradation of iron protected with polypyrrole films for biodegradable cardiovascular stents
The degradation of polypyrrole (PPy) coated iron is studied in phosphate buffer saline solution at 37 degrees C by odd random phase multisine electrochemical impedance spectroscopy (ORP-EIS). PPy is electro-polymerized with anti-inflammatory salicylates incorporated in the film, as a drug release system. The modelling of EIS over time provides the quantitative description of the corrosion behaviour of the material. Thus, the reliable analysis of the degradation stages of PPy coated iron is attained. The outcome of the present study shows that the degradation of iron can be tailored by tuning the properties of the PPy coating for possible medical applications. (C) 2017 Elsevier Ltd. All rights reserved.