Polymer Bulletin, Vol.39, No.1, 67-71, 1997
Improved Infrared Spectroscopic Method for the Analysis of Degree of N-Deacetylation of Chitosan
Various infrared spectroscopic techniques for the analysis of degree of N-deacetylation of chitosan were evaluated for accuracy. A new method was proposed which involved development of a calibration curve using the absorbance intensities of the chitosan infrared spectrum at 1655 and 3450 cm(-1). The, degree of N-deacetylation of various chitosan samples was then determined by using the absorption ratios [A(1655)/A(3450)] in the equation of the calibration curve. This method yielded results which were superior to those of any of the previously reported infrared spectroscopic methods tested herein.