Applied Microbiology and Biotechnology, Vol.45, No.6, 844-850, 1996
Microbial Sensors for Naphthalene Using Sphingomonas Sp B1 or Pseudomonas-Fluorescens Ww4
Amperometric biosensors for naphthalene were developed using either immobilized Sphingomonas sp. B1 or Pseudomonas fluorescens WW4 cells. The microorganisms were immobilized within a polyurethane-based hydrogel, which was used for a microbial biosensor for the first time, Both strains were shown to be equally suited for the quantification of naphthalene in aqueous solutions. The biosensors were tested in a now-through system and a stirred cell (batch method). In both systems a linear response down to the detection limit was obtained. Measurements in the flow-through system gave sensitivities of up to 1.2 nA mg(-1) l(-1) and a linear range from 0.03 mg/l to 2.0 mg/l. The response lime (t(95)) was 2 min and the sample throughput six per hour; the repeatability was within +/- 5%. With the batch method, sensitivities of between 3 nA mg(-1) l(-1) and 5 nA mg(-1) l(-1) and a linear range of 0.01-3.0 mg/l were obtained; the response time was between 3 min and 5 min. The sensors reached an operational lifetime of up to 20 days. The sensitivity of both sensors for naphthalene was, in most cases, more than four times higher than for various other substrates.