화학공학소재연구정보센터
Thermochimica Acta, Vol.229, 15-25, 1993
Experimental Investigations on Factors Influencing the Calibration of Heat-Flux Calorimeters
The influences of temperature, heating rate, and of the mass and heat capacity of the sample on the temperature and heat calibration, the time constant and the coefficient of thermal conduction are examined for a power-compensating, MCB-type DSC. The results for a metallic and an organic substance are compared.