Chemistry Letters, Vol.29, No.10, 1158-1159, 2000
Cherenkov counting efficiencies for beta(-)-emitters in dry state in glass vials
Cherenkov counting efficiencies for standardized beta (-)-emitters in the dry state at the centre of air-filled glass vials were measured with a liquid scintillation spectrometer. Cherenkov counting efficiencies, which are plotted as a function of the average energy of beta (-)-particles and the internal conversion electrons, give a straight line on log-log scale.