Electrophoresis, Vol.24, No.19-20, 3476-3483, 2003
Thousandfold signal increase using field-amplified sample stacking for on-chip electrophoresis
Field-amplified sample stacking (FASS) leverages conductivity gradients between a volume of injected sample and the background buffer to increase sample concentration. A major challenge in applying FASS to on-chip assays is the initial setup of high-conductivity gradient boundaries in the region of the injected sample volume. We have designed, fabricated, and characterized a novel FASS-capillary electrophoresis (CE) chip design that uses a photoinitiated porous polymer structure to facilitate sample injection and flow control for high-gradient FASS. This polymer structure provides a region of high flow resistance that allows the electromigration of sample ions. We have demonstrated an electropherogram signal increase by a factor of 1100 in electrophoretic separations of fluorescein and Bodipy with, respectively, 2 muM and 1 muM initial concentrations.
Keywords:capillary electrophoresis;field-amplified sample stacking;microfluidic chip;miniaturization;photoinitiated polymerization