화학공학소재연구정보센터
Electrophoresis, Vol.27, No.24, 5128-5131, 2006
Studying drug-plasma protein interactions by two-injector microchip electrophoresis frontal analysis
We developed a simple, rapid, and sensitive two-injector microchip electrophoresis frontal analysis (MCE-FA) method for studying drug-plasma protein interactions. In this method, large volumes of a reference sample and drug-plasma protein mixture were simultaneously introduced into the respective sections of the microchannel through the separated injectors and then electrophoresed. Since the reference sample did not meet with the interacting species during migration, it could be used as an external standard. The interaction between heparin and HSA was quantitatively characterized as a model system. The binding constant was found to be (1.53 +/- 0.01) X 10(4) M-1.