화학공학소재연구정보센터
Electrophoresis, Vol.29, No.8, 1612-1618, 2008
Automation of nanoflow liquid chromatography-tandem mass spectrometry for proteome and peptide profiling analysis by using a monolithic analytical capillary column
An automated nano-LC-MS/MS platform without trap column was established, which only used a 20 cm lauryl methacrylate-ethylene dimethacrylate (LMA-EDMA) monolithic capillary column to allow preconcentration and separation of peptides. The monolithic column had the advantages of good permeability and low backpressure resulting in higher flow rates for capillary columns. Tryptic digests of bovine albumin and yeast protein extract were tested using the monolithic column system. High proteomic coverage using this approach were demonstrated in this study. Furthermore, peptide samples extracted from mouse liver were separated by using the monolithic column system combined with size-exclusion chromatography prefractionation. This monolithic column system might be a promising alternative for the automated system previously using a trap column for routine proteome and peptide profiling analysis.