Applied Surface Science, Vol.255, No.1, 122-124, 2008
A novel design for a variable energy positron lifetime spectrometer
We present computer simulations of a new design of a variable energy positron lifetime beam that uses for a start signal the secondary electron emission from a 25-nm thick carbon foil (C-foil) located in front of the sample. A needle of similar to 30 mu m diameter is positioned on-axis right behind the foil, creating a radial electric field that deflects the secondary electrons radially outward so as to miss the sample and to hit the micro-channel plate (MCP) detector placed down beam. The MCP signal provides the start signal for the positron lifetime spectrometer. A grid can be further introduced between the sample holder and the MCP to yield a cleaner signal by preventing the positrons with large transmitted scattering angle from hitting the MCP. The cylindrical symmetry of this design reduces the experimental complexity and offers good timing resolution. We show that the design is robust against the transmitted energy and angle of the secondary electrons and positrons. (C) 2008 Elsevier B.V. All rights reserved.