화학공학소재연구정보센터
학회 한국화학공학회
학술대회 2004년 가을 (10/29 ~ 10/30, 호서대학교(아산캠퍼스))
권호 10권 2호, p.1998
발표분야 이동현상
제목 Design of Artificial Lung using the PZT Actuator
초록 The purpose of this work was to assess and quantify the beneficial effects of gas exchange according to the various frequencies of the sinusoidal wave that was excited by the PZT actuator, for patients suffering from acute respiratory distress syndrome (ARDS). The experimental design and procedure are then applied to the construction of the new device to be used to assess the effectiveness of the membrane vibrations. We developed an analytical solution for the hydrodynamics of flow through a bundle of the sinusoidal vibrated hollow fibers with the intention of gaining insight into how the wall vibrations might enhance the performance of the intravenous lung assist device. As a result, the vibration method is very effective in increase gas transport. The gas exchange efficiency for the vibrating intravascular lung assist device can be increased by emphasizing the following design features: consistent and reproducible fiber geometry, and most importantly, an active means of enhancing convective mixing of water around the hollow fiber membranes. The experimental design and procedure are then given for a device used to assess the effectiveness of membrane vibrations.
저자 김기범1, 김성종2, 정경락1, 권대규1
소속 1전북대, 2익산대
키워드 articial lung; gas transfer; pzt actuator
E-Mail
VOD VOD 보기
원문파일 초록 보기