화학공학소재연구정보센터
학회 한국화학공학회
학술대회 2010년 봄 (04/22 ~ 04/23, 대구 EXCO)
권호 16권 1호, p.397
발표분야 생물화공
제목 Detection of penicillin concentration using beta-lactamasecoated polyaniline biosensor
초록 Lactamase coated polyaniline biosensor for the detection of penicillin has been developed. The polyaniline was dispersed on the surface of carbon paste and lactamase was covalently attached to the polyaniline. The operational characteristics of developed biosensor were investigated using differential amplification method which are cyclic voltammetry and chronoamperometry.
The developed biosensor has been characterized using various concentrations of penicillin G. We used cyclic voltammetry to understand the electrochemical property of biosensors. The biosensor coated with polyaniline showed better hysteresis and linear range than that without polyaniline. The developed penicillin biosensor possesses a low detection limit of 100nM. The polyaniline biosensor showed linear and stable response in the range of 100 ~ 1000 nM penicillin concentration.
저자 박새로미, 이중헌
소속 조선대
키워드 β-lactamase (Penicillinase); Penicillin G sodium; Immobilization; polydimethylsiloxane; cyclicvoltametry; chronoamperometry
E-Mail
원문파일 초록 보기