화학공학소재연구정보센터
학회 한국화학공학회
학술대회 2002년 가을 (10/24 ~ 10/26, 서울대학교)
권호 8권 2호, p.3917
발표분야 생물화공
제목 LOC(랩온어칩) 응용을 위한 마이크로패턴에 형성된 전기화학식 면역 바이오센서의 신호검출 방법
초록 With the advent of LOC(Lab-on-a-chip) and μ-TAS(total analysis system) technology, there has been an increasing demand for efficient methodologies for DNA, protein, and metabolites analyses (biosensing). This presentation describes a strategy of signal registration from micropatterned immunosensors that converts antigen-antibody binding reaction into electrochemical signals. Of the transduction techniques in use such as electrochemical, optical, piezoelectric, etc., amperometric signaling was chosen for the relatively direct application of the microelectronics, which is advantageous for sensor miniaturization and automation. The cyclic voltammetric method, tracking biocatalyzed precipitation of insoluble products onto the sensing surface and subsequent decrement in the electrode area, was used for signal registration. As a model immunosensing with the micro-fabricated device, we have investigated the functionalization of ferritin (antigen) to the PAMAM dendrimer-assisted sensing surface and its biospecific interaction with anti-ferritin antiserum.
저자 윤현철, 양해식, 고종수, 김윤태
소속 한국전자통신(연)
키워드 biosensor; lab-on-a-chip; bioMEMS
E-Mail
VOD VOD 보기
원문파일 초록 보기